The TSD203 connects to the GSR100C amplifier for the purpose of skin conductance recording. Two Ag-AgCl, unpolarizable, electrodes are mounted in individual, ergonomically-designed, polyurethane housings for improved contact. The electrodes are attached to the fingers by Velcro® straps. The electrodes have a 6mm (dia) contact area with a 1.6mm cavity to accommodate electrode gel. The TSD203 incorporates shielded construction to minimize noise interference and improve recordings.
Usage Recommendations When using GEL101 it is important that the gel has a chance to be absorbed and make good contact before recording begins. Accordingly:
- Abrade the skin at the point of electrode contact; you can use a nail file or an EL PAD abrasion pad.
- Apply GEL101 to the skin at the point of electrode contact and rub it in.
- Fill the TSD203 electrode cavity with GEL101.
- Attach the TSD203 electrode to the subject.
- Wait 5 minutes (minimum) before starting to record data.
To verify the Gain setting of the GSR100C: - Calibrate AcqKnowledge for lower frequency response at DC: In the Scaling window, set the Input voltages so they map to the DC conductance ranges indicated by the sensitivity setting. For example, if the GSR100C is set to a Gain of 5 µmho/V, then 0V will map to 0 µmhos or infinite resistance and 1V will map to 5µmho or 200kohm.
- Place the lower frequency response to DC.
- Set the Gain switch on the GSR100C to 5µmho/V.
- Perform measurement with electrodes disconnected.
· AcqKnowledge should produce a reading of 0 µmho. - Insulate a 100kohm resistor and place it from electrode pad to electrode pad (resistor must be insulated from fingers).
- Perform measurement with electrode-resistor setup.
· AcqKnowledge should produce a reading of 10 µmho.
TSD203 Specifications
Electrode Type: Ag/AgCl (unpolarizable)
Surface Area: 6mm contact area
Dimensions: 16mm (long) x 17 mm (wide) x 8mm (high) (each electrode)
Weight: 4.5 grams (each electrode)
MRI Compatible: Yes (no ferrous parts)
Sterilizable: Yes (see Biopac for details)
Cable Length: 3 meters (shielded)
Interface: GSR100C
TEL100C Compatibility: SS3A
Cleaning the TSD203:
- Do not leave GEL in the cavity after use. The electrode cavity must be left clean and dry. If GEL is left in the cavity, the Ag-AgCl electrode disk could degrade quickly with time because the electrode surface is somewhat porous to promote good conductivity to the GEL.
- To clean the reusable TSD203, use a cotton swab or toothbrush with tap water.
- Use any lab cleaner with pumice (such as Ajax) with a cotton swab or toothbrush to remove any dark residue from the electrode surface.
- Use Hydrogen Peroxide solution (2-3%) to brighten the electrode surface (optional) or to sterilize the electrode. Do not place the electrode in solution, but simply clean the electrode surface using a cotton swab. Dry the electrode off completely before storage.
|
|