The TSD160 series differential pressure transducers are designed for low range pressure monitoring. The transducers plug directly into the DA100C general-purpose differential amplifier. The differential pressure ports are located on the front of the transducers and are easily connected to breathing circuits, pneumotachs or plethysmograph boxes. These transducers are very useful for interfacing a variety of small animal pneumotachs or plethysmographs to the MP System. The transducers are extremely sensitive and come in three ranges to suit a number of different applications. RX137 flow heads connect to the TSD160B differential pressure transducer via standard 4mm ID tubing.
TSD160B Specifications
Operational Pressure: ±12.5cm
H2O Overpressure (max): ±375cm H2O
Voltage Output (normalized to 1 volt excitation): 131µV/cm H2O
Warm-up Drift: ±50µV
Stability: ±100µV
Combined Linearity and Hysteresis Error: ±0.05%
Dynamic Response: 100Hz
Connection Ports: Accepts 3mm to 4.5mm ID tubing
Dimensions: 8.3cm (high) x 3.8cm (wide) x 3.2cm (deep)
Weight: 76 grams
Operating Temperature: 0 to +50 °C (compensated)
Storage Temperature: -40 to +125 °C
Interface: DA100C
|
|