The TSD125 series force transducers interface with the DA100C general-purpose transducer amplifier. The TSD125 series have been designed for in-vitro tissue bath studies, in-vivo force measurements and other applications where low noise, accuracy and repeatability are critical. The most sensitive units are ideal for use with small preparations such as aortic rings. The TSD125 series utilizes a precision strain gauge element, which has superb low noise characteristics and excellent linearity. The transducers come with a versatile mounting rod assembly to support a variety of orientations.When fine tension and position adjustments are required, clamp the mounting rod to an HDW100A Force Transducer Tension Adjuster (not included).
The following table shows the Full Scale Range (FSR) of the TSD125 Series Precision Fixed Range Force Transducers, and typical noise levels with 10Hz and 1Hz Low Pass Filters.
| Device |
FSR |
10Hz |
1 Hz |
| TSD125C |
50 gram |
2.5 mg |
1 mg |
| TSD125D |
100 gram |
5 mg |
2 mg |
| TSD125E |
200 gram |
10 mg |
4 mg |
| TSD125F |
500 gram |
25 mg |
10 mg |
(See also the TSD105A Variable Range Force Transducer.)
TSD125E Specifications
Full Scale Output: 1mV/V (normalized to 1V excitation)
Full Scale Range: 200 gram
Noise, 10 Hz Bandwidth: 10 mg
Noise, 1 Hz Bandwidth: 4 mg
Nonlinearity: <±0.025% FSR
Hysteresis: <±0.05% FSR
Nonrepeatability: <±0.05% FSR
30 Minute Creep: <±0.05% FSR
Temperature Range: -10°C to 70°C
Thermal Zero Shift: <±0.03% FSR/°C
Thermal Range Shift: <0.03% Reading/°C
Recommended Excitation: 10 VDC (max)
Dimensions: 100mm (long) x 19mm (wide) x 25mm (thick)
Weight: 250 grams
Mounting Rod: 9.5mm (dia)- variable orientation
Cable Length: 3 meters
Interface: DA100C